Probes and Biopsy  13-17 
13.1.8  Storage and Transportation for Sensor Support 
(Fusion) 
When all examinations for the day have been completed, confirm that the sensor support is in good 
condition. After disinfecting the sensor support, confirm that the sensor support is in good condition and 
stored in a suitable place. 
1.  To prevent the sensor support from being damaged, DO NOT store it in locations where it may be 
exposed to: 
  Direct sunlight 
  Sudden changes in temperature 
  Dust 
  Excessive vibration 
  Heat generators 
2.  Store and transport sensor support under following ambient conditions: 
  Ambient temperature: -20°C ~ 55°C 
  Relative humidity: 20% ~95% (no condensation) 
  Atmospheric pressure: 700 hPa ~1060 hPa 
13.1.9  Sensor (Fusion Imaging) Cleaning and Disinfection 
CAUTION: 
1.  When performing cleaning and 
disinfection of the sensor to prevent 
infection, wear sterile gloves. 
After disinfection, rinse the sensor thoroughly with sterile water
 
to remove all chemical residues. Chemical residues may be 
harmful to the human body. 
3
No cleaning and disinfecting may result in the sensor becoming 
a source of infection. 
 
4 
The efficacy of disinfectants and sterilizing solutions is not 
guaranteed by MINDRAY. Contact the manufacturers for 
information on the activity of the products. 
 
Disinfecting with Sprays or Wipes 
CAUTION: 
Use protective eyewear when disinfecting using sprays. 
1.  Wear sterile gloves to prevent infection. 
2.  Spray or wipe the sensor (fusion) with a disinfectant. Follow the disinfectant manufacturer’s 
recommended contact time and mode. 
3.  Remove any residue with a water-moistened soft cloth on the sensor. 
4.  Wipe off water on the sensor using sterile cloth or gauze after washing.