13-34 Probes and Biopsy
4. Hold the transducer. Press the biopsy needle to separate needle guided V-shaped block from
pressure position of the needle.
5. Put the needle into the needle guided-bracket, and the needle leans to V-shaped block.
6. Hold the transducer, and release the pressure position of the needle. Adjust the needle-type
adjusting nut manually (following the direction of the arrow). The needle moves smoothly at the
vertical direction due to its gravity.
NGB-039
(1) Put the sterile sheath on the probe.
(2) Select a proper needle-guided bracket, and match the locating groove with the tab of the
transducer. Mount the bracket onto the transducer.
(3) Tighten the pinch nut of the needle-guided bracket to confirm that the needle-guided bracket is
properly installed on the probe. Loosen the angle adjusting nut.