Probes and Biopsy  13-55 
5.  Tighten the knob, and make sure not to break the sterile sheath.   
 
6.  Turn on the magnetic navigator power and perform needle navigation guiding operation after the signal 
is stabilized. Refer to “5.17.3 Screen Description” for details. 
7.  Perform needle navigation guiding; refer to “13.3.1 Basic Procedures for Needle Navigation 
Guiding” for details. 
8.  Anticlockwise eTRAX needle interface to take out the needle from trocar after the eTRAX needle 
access the target tissue. 
 
9.  Select the proper Biopsy/melt needle according to the actual situation. 
 
13.3.3  Accuracy Verification 
WARNING: 
This manual describes simple accuracy verification only. 
Perform elaborate accuracy verification procedure based 
on clinical scenario if necessary. 
If the needle is bent, projective line of the needle and 
projective point of the tip will not suit the actual situation. 
Be sure to conduct accuracy verification before needle 
navigation. Perform needle navigation when the measured 
result is in the acceptable range.
 
 
Accuracy error is measured by comparing actual biopsy depth and the measured depth in the ultrasound 
image. Needle navigation guidance feature should be activated only when the error is acceptable. 
1.  Connect all accessories of magnetic navigator. Refer to “13.3.2 Installation” for details. 
2.  Precisely measure the distance between tip to the bottom. 
Anticlockwise interface