EasyManuals Logo

Mindray Resona R9 Exp Service Manual

Mindray Resona R9 Exp
365 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #229 background imageLoading...
Page #229 background image
Structure and Assembly/Disassembly 9-67
2. Unplug 6 cables from the AC power assembly. Unscrew 4 M4 X8 cross panhead screws and
one M4 X 8 panhead screw from the AC power assembly to remove the AC power assembly
(FRU).
9.3.31.5 Transformer
1. See Chapter 9.3.31 for details.
2. Unscrew one M4 X 8 cross panhead screw and 4 M4 X 8 cross panhead screws from the
grounding terminal of the transformer with the screwdriver (M3, M4) to remove the transformer
(FRU).
AC power assembly
(
FRU)

Table of Contents

Other manuals for Mindray Resona R9 Exp

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Resona R9 Exp and is the answer not in the manual?

Mindray Resona R9 Exp Specifications

General IconGeneral
BrandMindray
ModelResona R9 Exp
CategoryDiagnostic Equipment
LanguageEnglish

Related product manuals