Product Specifications
3-7
SpO
2
(%) 1%
PR (bpm) 1
Response time
Under the condition that the PR is 75BPM and the mean time is 8s, the maximum
response time for the SpO
2
value to increase from 60% to 95% is 20s.
3.2.9.2 9006 SpO2
Range: 0 - 100%
Resolution: 1%
Accuracy
No motion conditions:
SpO
2
range Patient Type
70% to 100% 0 - 69%
Adult (including pediatric) ± 2 digits Undefined
Neonate ± 3 digits Undefined
Updating cycle of SpO
2
values: 1s
PR
Range: 20 - 254 BPM
Resolution: 1bpm
Accuracy
Patient Type PR range Accuracy( No motion conditions)
Adult/Pediatric/Neonate 20-254 bpm ± 3 digits
3.2.9.3 Nellcor SpO
2
Range and accuracy for
SpO
2
measurement:
Sensor Accuracy
1
MAX-A, MAX-AL, MAX-N, MAX-P,
MAX-I and MAX-FAST
70% - 100% ± 2 %
0% - 69% Undefined
OxiCliq A, OxiCliq N, OxiCliq P and
OxiCliq I
70% - 100% ±2.5 %
0% - 69% Undefined
D-YS, DS-100A, OXI-A/N and
OXI-P/I
70% - 100% ± 3 %
0% - 69% Undefined
MAX-R, D-YSE and D-YSPD 70% - 100% ± 3.5 %
0% - 69% Undefined