Machine test and Material List
5-7
5.3 NIBP Calibration Method
Figure 1 NIBP Static Pressure Calibration Method
Calibration method: Increase the pressure at intervals of 50mmHg(6.7kPa). The maximum
difference between the monitor and calibrator at any pressure point within the measurement
range does not exceed ±3mmHg (±0.4kPa). Similarly, when decreasing the pressure, the
difference between the monitor and calibrator at any pressure point within the measurement
range should not exceed±3mmHg (±0.4kPa).