Machine test and Material List 
5-2 
 
5.1.4  TEMP Test 
The monitor shall correctly measure normal body temperature. When monitoring ambient 
temperature in MONITOR mode, the temperature reading displayed shall be in normal range and 
continuously shown on the temperature display area.   
 
5.1.5 Recorder Print Test  
1.  Print SpO2 graph. The recorder should print correctly and the printed results should 
be clear and consistent. If set such faults such out of paper, etc., corresponding 
prompts should be given. After the fault is removed, the VS-800 should be able to 
work correctly. 
2.  Print trend data. The recorder should print in accordance with what described in the 
Operation Manual. 
5.1.6 Clock Test  
Verify the accuracy of clock counting when conducting the system test, and then set the clock to the 
current time. 
5.1.7 Whole Parameter Test  
Plug all monitoring parameters and monitor them one by one. During the simultaneous monitoring of all 
parameters, the parameters should not interfere with each other. 
5.1.8 Set Defaults at the time of loading Software  
Perform such main menu operations as patient information management, trend data review, 
system setup, etc. The system responds correctly and no obvious error occurs during 
functioning. Each function meets the product requirements.