EasyManua.ls Logo

Mindray VS-800 - Page 62

Mindray VS-800
74 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Machine test and Material List
5-2
5.1.4 TEMP Test
The monitor shall correctly measure normal body temperature. When monitoring ambient
temperature in MONITOR mode, the temperature reading displayed shall be in normal range and
continuously shown on the temperature display area.
5.1.5 Recorder Print Test
1. Print SpO2 graph. The recorder should print correctly and the printed results should
be clear and consistent. If set such faults such out of paper, etc., corresponding
prompts should be given. After the fault is removed, the VS-800 should be able to
work correctly.
2. Print trend data. The recorder should print in accordance with what described in the
Operation Manual.
5.1.6 Clock Test
Verify the accuracy of clock counting when conducting the system test, and then set the clock to the
current time.
5.1.7 Whole Parameter Test
Plug all monitoring parameters and monitor them one by one. During the simultaneous monitoring of all
parameters, the parameters should not interfere with each other.
5.1.8 Set Defaults at the time of loading Software
Perform such main menu operations as patient information management, trend data review,
system setup, etc. The system responds correctly and no obvious error occurs during
functioning. Each function meets the product requirements.

Related product manuals