5. Analog output
5−1. Voltage output
1
Output
Output ±10 V at the time of the rated output of torque transducer.
The analog output range can be changed by attenuator switch (DIP401).
DIP401
Output range
only 1 is ON
0±10 V
only 2 is ON
0±5V
only 3 is ON
0±2V
only 4 is ON
0±1V
Please make the setting of switch as only one.
2
Electrical condition
D Load resistance :2kΩ or more
D Capacitance load :0.1 μF or less
5−2. Current output (option)
Adjustment can be made when the optoinal current output is set.
Please execute the setting after turning off the power.
1
Output
The following setting of the current output can be made at the rated output from the torque
transducer.
i) When the output of 12.00 mA±8.00 mA at 0〜±rated output of torque
The current output adjustment switch (SW402) is set to the upper part.
ii) When the output of 4.00 mA〜20.00 mA at 0〜+rated output of torque.
The current output adjustment switch (SW402) is set to the lower part.
The main setting of the voltage output and the current output is as follows;
a ) When 12 mA of the current output at 0 V of the voltage output
Voltage output
Current output SW402 GAIN
−10 V〜10 V 4mA〜20 mA Upper side 1
−5V〜5V 4mA〜20 mA Upper side 2
−2V〜2V 4mA〜20 mA Upper side 5
−1V〜1V 4mA〜20 mA Upper side 8
b ) When 4 mA of the current output at 0 V of the voltage output
Voltage output
Current output SW402 GAIN
0V〜10 V 4mA〜20 mA Lower side 2
0V〜5V 4mA〜20 mA Lower side 4
0V〜2V 4mA〜20 mA Lower side 8
0V〜1V 4mA〜20 mA Lower side F
2
Electrical condition
D Load resistance :510 Ω or more