2 INSTALLING THE LINEAR GAGE
2
INSTALLING THE LINEAR GAGE
Insert the stem into the hole of the gage mount or stand and clamp it with the screw.
See the Linear Gage mount examples given below.
CAUTION
Do not clamp the Linear Gage using excessive force. This may cause the spindle to stick.
Do not mount the Linear Gage by clamping with a screw on the stem.
Do not mount the Linear Gage at any portion other than the stem.
Mount the Linear Gage perpendicular to the surface to be measured. Otherwise, a measurement error will
be produced as shown below.
Error = I
0
- I
1
Ensure that the gage body is not subjected to a force via the pull of the cable.
2.1
CONNECTING THE AIR HOSE (FOR 542-223, 542-230)
Connect the air hose as specified below :
Air pressure : 0.3-0.4MPa(3-4kgf/cm2)
Lubrication oil : Class 1 turbine oil(ISO VG32)
A
Air cylinder
Valve
Lubricator
Regulator
Air filter
Air regulator set
CAUTION
Never hold the air cylinder when installing the gage. This may damage portion A. Be sure that portion A
is not subjected to excessive force when connecting the air hose.