902B Abs Piezo Operation manual
25
Dual analog output
The 902B is available with dual analog output which can be used to provide an alternative output for
amplification of range or to emulate another transducer type while still using the MKS standard output.
This feature is a hardware option and must be specially ordered. Refer to part number specifications page 3.
The secondary analog output provides 12 bit resolution.
Command: @253AO2!15;FF
Command values: 10 to 319 (xy)
First digit (x) Use same parameters as primary analog output
Second digit (y) Use same parameters as primary analog output
Command reply: @253ACK105;FF
Query: @253AO2?;FF
Query reply: @253ACK105;FF
Factory default: 35