Operating Temperature is out of
the range
Use the thermometer in the range
of operating conditions
The sensor temperature has not
been stabilised
Wait 10 seconds and take a
measurement again.
The measurement is not accurate
or there is doubt of measurement
results.
Check to see if the infrared sensor
is clean or not.
Clean the infrared senosr with
cotton swab according to the user
manual.
Determine whether your
measuring method is correct or
not.
Please review the user manual
and ensure you are operating the
thermometer correctly.
Assess whether or not you hae
allowed the thermeter and the
patient to become stabilized in the
room for 30 minutes.
Please ensure the thermometer
and patient remain in the location
for at least 30 minutes before
recording a temperature.
Are you using the thermometer
indoors or outdoors.
Always do the measurements
indoors.
Determine if you might have held
the thermometer in your hand too
long which may affect its accuracy.
Place the thermometer in the
room where the measurement is
due to be taken, let it stabilize for
30 minutes before use.
Check to see if the Low Battery
icon is showing on the LCD screen.
Immediately replace the batteries
with new ones.
Product Safety & Warnings
This product was designed to measure human body temperature on the forehead, do not
use this for any other purpose.
Do not use this thermometer if it has been malfunctioning or it has been broken or
damaged.
This product is not waterproof. Do not use this product under any wet conditions and do
not immerse in any liquid or water.
This product should not be left alone with children or pets with not in use by an adult.
Do not store the thermometer in extreme temperatures below (-4°F / -20°C) or over
(131°F/ 55°C).
Do not use the product in a wet ear canal after swimming or bathing. It is possible to
damage the ear canal.