| 37 |
(1)The equipment complies with FCC RF exposure limits set forth
for an uncontrolled environment.
(2)The equipment must not be co-located or operation in
conjunction with any other antenna or transmitter.
FCC Label Compliance Statement:
This device complies with Part 15 of the FCC Rules. Operation is
subject to the following two conditions:
(1)this device may not cause harmful interference, and
(2)this device must accept any interference received, including
interference that may cause undesired operation.
MOCAarm essential performance per IEC 80601-2-30 additional
essential performance requirements:
201.12.1.102 Limits of the error of the manometer from
environmental conditions
Over the temperature range of 5 °C to 40 °C (41 °F ~ 104 °F ) and
the relative humidity range of 15 % to 93 %(non-condensing), the
maximum error for the measurement of the CUFF pressure at any
point of the NOMINAL measurement range shall be less than or
equal to ± 3 mmHg (± 0.4 kPa) or 2 % of the reading, whichever is
greater.
201.12.1.107 Reproducibility of the blood pressure determination
The laboratory Reproducibility of the BLOOD PRESSURE
DETERMINATION of the AUTOMATED SPHYGMOMANOMETER shallbe
less than 3 mmHg (0.4 kPa).