Guidance and manufacturer’s declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environments
listed below, and should only be used in such environments:
IMMUNITY
TEST
IEC 60601
TEST LEVEL
COMPLIANCE
LEVEL
ELECTROMAGNETIC
ENVIRONMENT –
GUIDANCE
Electrostatic
discharge
(ESD)
IEC 61000-4-2
± 8 kV contact
discharge
± 15 kV air discharge
± 8 kV contact
discharge
± 15 kV air
discharge
In the case of air discharge
testing, the climatic conditions
shall be within the following
ranges:
Ambient Temperature:15°C~35°C,
Relative Humidity:30%~60%.
Power
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
30 A/m
50 or 60 Hz
30 A/m
50 or 60 Hz
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital environment.
Electrical fast
transient/
burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for input/
output lines
± 2 kV for power
supply lines
± 1 kV for input/
output lines
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
AC Power port
±1 KV Line to Line
AC Power port
±1 KV Line to Line
Mains power quality should be
that of a typical commercial or
hospital environment.
interruptions
and voltage
variations
on power
supply
input lines
IEC 61000-4-11
0% UT; 0.5 cycle
At 0°,45°,90°,135°,180°,
225°,270°and 315°.
0 % UT; 1 cycles
70 % UT; 25/30 cycles
0 % UT; 250/300 cycle
0% UT; 0.5 cycle
At 0°,45°,90°,135°,
180°,225°,270°and
315°.
0 % UT; 1 cycles
70 % UT; 25 cycles
0 % UT; 250 cycle
Mains power quality should be
that of a typical commercial
or hospital environment. If the
user of the device requires
continued operation during
power mains interruptions,
it is recommended that the
device be powered from an
uninterruptible power supply or
a battery.
APPENDIX