| 42 |
APPENDIX
Test specifications for enclosure port immunity to RF
wireless communications equipment.
Test frequency
(MHz)
Modulation
IMMUNITY TEST LEVEL
(V/m)
385 Pulse modulation 18 Hz 27
450 FM ± 5 kHz deviation 1kHz sine 28
710
Pulse modulation 217 Hz 9745
780
810
Pulse modulation 18 Hz 28870
930
1720
Pulse modulation 217 Hz 281845
1970
2450 Pulse modulation 217 Hz 28
5240
Pulse modulation 217 Hz 95500
5785
NOTE:
If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the
transmitting antenna and the ME EQUIPMENT or ME SYSTEM may be reduced to 1 m,
the 1 m test distance is permitted by IEC 61000-4-3.
a). The carrier shall be modulated using a 50% duty cycle square wave signal.
b). AS an alternative to FM modulation, 50% pulse modulation at 18 Hz may be used
because while it does not represent actual modulation, it would be worst case.