A MULTI-DETECTION MICROPLATE READER WITH TWO-MODE CUVETTE PORT
SpectraMax
®
M2/M2
e
user guide

Questions and Answers

  • C
    Christine BradfordSep 8, 2025
    What does 'NVRAM CRC corrupt' mean on Molecular Devices Laboratory Equipment?
    • C
      Carol WaltersSep 9, 2025
      An 'NVRAM CRC corrupt' error in Molecular Devices Laboratory Equipment means that the CRC for the NVRAM data is corrupt.
  • W
    William RiversAug 25, 2025
    What to do if Molecular Devices SpectraMax M2e shows signal level saturation?
    • M
      Mary ChavezAug 25, 2025
      If the signal level is saturated on your Molecular Devices Laboratory Equipment, verify that the cuvette door is not open.
  • C
    Chelsea CarpenterAug 25, 2025
    What to do if Molecular Devices SpectraMax M2e shows reference level saturation?
    • M
      Michael ChambersAug 26, 2025
      If you are getting reference level saturation on your Molecular Devices Laboratory Equipment, a possible solution is to check if the cuvette door is open.
  • T
    tnorrisSep 22, 2025
    What does 'assert failed' mean on Molecular Devices SpectraMax M2e?
    • J
      Jacqueline HatfieldSep 22, 2025
      Firmware error.
  • J
    jessica51Sep 9, 2025
    What does 'input line too long' mean for Molecular Devices SpectraMax M2e?
    • N
      Nancy BennettSep 9, 2025
      The 'input line too long' error in Molecular Devices Laboratory Equipment means there are too many characters in the input line.
  • R
    Rhonda MartinezSep 14, 2025
    What does 'NVRAM Carriage offset error' mean on Molecular Devices SpectraMax M2e Laboratory Equipment?
    • M
      Melissa RothSep 14, 2025
      An 'NVRAM Carriage offset error' in Molecular Devices Laboratory Equipment means the carriage offset data is unreasonable.
  • J
    Jennifer BrownSep 3, 2025
    What is a carriage motion error in Molecular Devices SpectraMax M2e Laboratory Equipment?
    • S
      Suzanne CombsSep 4, 2025
      A 'carriage motion error' in Molecular Devices Laboratory Equipment means the carriage did not move to either of its photo interrupts in a reasonable time, or cannot find its photo interrupt.
  • S
    Sean AshleySep 4, 2025
    What does 'plate air cal fail, low light' mean on Molecular Devices SpectraMax M2e?
    • B
      Brad MooreSep 4, 2025
      A 'plate air cal fail, low light' error in Molecular Devices Laboratory Equipment means the minimum signal/reference ratio was not met during air calibration.
  • A
    Amanda SmithSep 23, 2025
    What does 'unknown interrupt' mean on Molecular Devices Laboratory Equipment?
    • A
      Anthony McclureSep 23, 2025
      Firmware error.
  • L
    Lonnie BurchAug 29, 2025
    What causes a 'thermistor faulty' error in Molecular Devices Laboratory Equipment?
    • E
      Emily MyersAug 29, 2025
      A 'thermistor faulty' error in Molecular Devices Laboratory Equipment means the thermistor is either faulty or disconnected, there's a main board problem, or the ambient temperature is out of range. The equipment is unable to read a reasonable thermistor value.

Summary

2. Principles of Operation

2.4 PathCheck and Correction Methods

Details on pathlength compensation technology and its correction methods.

3. Installation Procedures

3.1 Unpacking

Instructions for safely removing the instrument from its packaging.

3.2 Setting Up the Instrument

Guidelines for placing, connecting, and initial setup of the instrument.

4. Operation Guide

4.3 Preparing for Reading

Steps for powering on, setting temperature, and loading samples.

4.6 Optimizing Fluorescence Assays

Guidance on adjusting settings for optimal fluorescence detection.

5. Maintenance and Care

5.6 Changing the Fuses

Step-by-step instructions for replacing instrument fuses.

6. Troubleshooting

6.1 Opening the Drawer Manually

Procedure for manually opening the drawer in case of errors.

6.2 Error Codes and Probable Causes

Lists error codes and their likely causes and remedies.

7. Specifications

Molecular Devices SpectraMax M2e Specifications