6
5. Setting up the device
Menu level Parameterization level
Setting the decimal point, :
Default:
The decimal point on the display can be moved with [▲] [▼] and confirmed with
[P]. The display then switches back to the menu level again.
Setting the display time, :
Default:
The display time is set with [▲] [▼]. The display moves up in increments of 0.1 up
to 1 second and in increments of 1.0 to 10.0 seconds. Confirm the selection by
pressing the [P] button. The display then switches back to the menu level again.
5.3. Programming interlock
Activation/deactivation of the programming lock and completion of the
standard parameterization,
Default:
Choose between the deactivated key lock (works setting) and the activated
key lock with the aid of the [▲] [▼] keys. Make the selection with [P]. After
this, the display confirms the settings with " , and automatically switches to
operating mode. If was selected, the keyboard is locked. To get back into the
menu level, press [P] for 3 seconds in operating mode. Now enter the (works
setting ) that appears using the [▲] [▼] keys plus [P] to unlock the keyboard.
appears if the input is wrong.
dann