1.3 Design
The design of the machine has one purpose: low uncertainty. The force the calibrang machine produces is
transferred directly to the unit under test without any machine loss or mechanical interference. Fricon be-
tween the reference standard and the unit under test has been eliminated. This is essenal because even slight
fricon, which is inherent in other types of calibrang machines, frequently results in gross errors. Non-axial
loading, another potenal error source, is eliminated by the design of the machine and its accessories.
1.4 Reference Standard
While a reference standard, such as a load cell, is essenal to calibraon, it is not supplied as part of the
Universal Calibrang Machine. The reference standard is always loaded in compression whether performing a
tension or compression calibraon.
Figure 2: Morehouse Ultra-Precision Load Cell