4.3 Tension Setup
Preparaon for a tension calibraon is similar to compression. The principle dierence is the manner in which
the unit under test is mounted in the machine. For a tension calibraon, the unit under test is connected be-
tween the lower yoke platen and the lower machine platen by means of tension members (pulling rods).
Tension members supplied with the unit under test may suce for this calibraon. The Morehouse Quick-
Change Tension Members can save me by making connecons easier and more ecient. Faster tesng and
calibraon turn-around can improve the boom line of your operaon. With this system, a set of self-aligning
tension members are installed in the tension side of the Universal Calibrang Machine. Then, a pair of Tension
Member Adapters, which are included in the kit can be used to quickly mount the unit under test into the ma-
chine and prepare for calibraon. Figure 21 illustrates how various adapters can be used for dierent instru-
ments with the same set of tension members.