98
Requirements for Computers or other Devices Connected to the Computers
1. The Veraviewepocs has been tested and found to comply with the limits for medical devices to the IEC
60601-1-2:2014 for electromagnetic disturbances. These limits are designed to provide reasonable protection
against harmful interference in a typical medical installation. This equipment generates uses and can radiate
radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful
interference to other devices in the vicinity. However, there is no guarantee that interference will not occur in a
particular installation. If this equipment does cause harmful interference to other devices, which can be
determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one
or more of the following measures:
- Reorient or relocate the receiving device.
- Increase the separation between the equipment.
- Connect the equipment into an outlet on a circuit different from that to which the other device(s) are
connected.
- Consult the nearest J. MORITA OFFICE, its representative or its dealer for help.
2. The following equipment connected to the analog and digital interfaces must be certified according to the
respective IEC standards (i.e. IEC 60950-1 or IEC 62368-1 for data processing equipment and IEC 60601-1
for medical equipment). Everybody who connects additional equipment to the signal input part or signal
output part configures a medical system, and is therefore responsible that the system complies with the
requirements of IEC 60601-1. If in doubt, consult the nearest J. MORITA OFFICE, its representative or its
dealer for help.
* Some of the following devices may cause some technical problems with the Veraviewepocs. Ask your nearest
J. MORITA OFFICE for proper selection of equipment and connections.
The following devices may not be located in the X-ray protection area or the patient vicinity except the
Hub if the Hub is conformed with IEC60601-1, IEC60950-1, or IEC 62368-1 and enclosure leakage
current is conformed with IEC 60601-1.
* The patient vicinity is the area where intentional or unintentional contact can occur between a patient or a
patient’s attendant and the above devices, or between a patient or a patient’s attendant and other persons
touching the above devices. This area extends 1.83 m beyond the perimeter of the bed (examination table,
dental chair, treatment booth, and the like) in its intended location, and vertically 2.29 m above the floor.
Connect only items that have been specified as part of medical electric system or specified as being
compatible with medical electric system.
Do not use multiple portable socket outlet nor extension cord for the system power supply.
* Computers or any other external devices must be connected in accordance with IEC 60601-1.
* Computers or any other external devices must be cleaned in accordance with the manufacturer’s instructions.
* Computers or any other external devices must be transport, storage, and operation in accordance with the
manufacturers’ instructions.