Guidance and Manufacturer’s Declaration Electromagnetic Immunity
This device is intended for use in the electromagnetic environment specified below.
The customer or the user of this device should assure that it is used in such an environment.
Immunity Test IEC 60601
Test Level
Compliance Level Electromagnetic Environment
Guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 V
ISM
(c)
/ amateur radio
frequency band: 6 V
150 kHz to 80 MHz
3 V/m
80 MHz to 2.7 GHz
27 V/m
385 MHz
28 V/m
450 MHz
9 V/m
710, 745, 780 MHz
28 V/m
810, 870, 930, MHz
28 V/m
1720, 1845, 1970 MHz
28 V/m
2450 MHz
9 V/m
5240, 5500, 5785 MHz
3 V
ISM
(c)
/ amateur radio
frequency band: 6 V
150 kHz to 80 MHz
3 V/m
80 MHz to 2.7 GHz
27 V/m
385 MHz
28 V/m
450 MHz
9 V/m
710, 745, 780 MHz
28 V/m
810, 870, 930, MHz
28 V/m
1720, 1845, 1970 MHz
28 V/m
2450 MHz
9 V/m
5240, 5500, 5785 MHz
Portable and mobile RF
communications equipment should be
used no closer to any part of this
device, including cables, than the
recommended separation distance
calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation distances
d 1.2 √P 150 kHz to 80 MHz
d 1.2 √P 80 MHz to 800 MHz
d 2.3 √P 800 MHz to 2.7 GHz
d √P Portable wireless RF
communication
equipment
Where P is the maximum output
power rating of the transmitter in watts
(W) according to the transmitter
manufacturer, E is the compliance
level in V/m and d is the
recommended separation distance in
meters (m).
Field strengths from field RF
transmitters, as determined by an
electromagnetic site survey
(a)
, should
be less than the compliance level in
each frequency range
(b)
.
Interference may occur in the vicinity
of equipment marked with the
following symbol:
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
(a)
Field strengths from fixed transmitters, such as base stations for ratio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicated theoretically
with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field strength in the location in which this device is used exceeds
the applicable RF compliance level above, this device should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as reorienting of relocating this device.
(b)
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
(c)
The ISM (Industrial, Scientific and Medical) bands between 0.15 MHz and 80 MHz are 6.765 MHz to 6.795
MHz; 13.553 MHz to 13.567 MHz; 26.957 MHz to 27.283 MHz; and 40.66 MHz to 40.70 MHz.
6
E

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Morita veraviewepocs 2d and is the answer not in the manual?

Morita veraviewepocs 2d Specifications

General IconGeneral
BrandMorita
Modelveraviewepocs 2d
CategoryMedical Equipment
LanguageEnglish

Related product manuals