Note for Force & Load Sensors
When performing a Load Cell Calibration, if choosing a 2 point calibration, load at 10% and 100%. If performing a 3 point
calibration and load at load at 10%, 50% and 100%.
The calibration program will guide you through the calibration by text in the information box. By selection in the
Calibration Type you can select Compression Only, Tension Only or Both if the Load cell supports this. Type in the
appropriate Full Scale value and select the units desired.