Dolito Bedienungsanleitung / Operators Manual
28
Technical Data
The patient is the intended operator for this unit. This device is quality controlled and fullls all
legal requirements for medical products. Therefore this device carries the CE-sign „CE 0123“.
The serial number is on the rear of the unit.
Manufacturer: MTR+ Vertriebs GmbH,
Kamenzer Damm 78, 12249 Berlin, Germany
www.mtrplus.com
Dimensions of the device: H 11,7 cm / W 6,6 cm / D 3,3 cm
Weight of the unit: 150 g (incl. battery)
Firmware version: 3.3 or later
Max. environmental conditions for transportation: -25 to +70 degree centigrade, 0-93%
atmospheric humidity. Recommended environmental conditions for storage: -25 to +40 degree
centigrade, 0-93% atmospheric humidity. The device shall not to be stored in direct sunlight.
Allow up to one hour for the device to return to normal operating temperature if stored in
conditions below +5°C or above +40°C.
Environmental conditions for use: +5 to +40 degree centigrade, 0-93% atmospheric humidity,
air pressure 700-1060 hPa. During use the surface temperature of the device can be 41°C or
higher, but never exceeds 48°C.
The Dolito is protected against debris larger than 12.5 mm in diameter from entering the
cabinet, is not water resistant and rated IP20.
Channels: Two isolated independent channels
Waveform: Asymmetrical, rectangular. bi-phasic with zero DC current
Intensity: 0 - 80 mA into 500 ohm load
80 mA maximum into 1000 Ohm load
70 mA maximum into 1500 Ohm load
(Indication only, actual intensity varies depending on
condition of skin electrodes)
Classication: Internally powered, constant current, continuous operation,
Type BF applied part, class IIa, IP20
Frequency: 1 - 150 Hz (Burst 35 - 150 Hz)
Pulse Width: 50 - 350 µS (Burst 50 - 200 µS)
Time Settings: Variable 1 minute to 9 hours 59 minutes (custom program)
Max. Voltage: 180 V without load (70V with 1000 ohm load)
Battery; 9 Volt 6LR61 / 6AM6 recommended