Dolito Bedienungsanleitung / Operators Manual
39
Programs
Abbreviations:
AdJ: Adjustment of intensity
AL: Alternate function of the channels A and B
BURST: The burst mode delivers 2 blocks of 9 impulses each per second
CONT: In the continuous mode a permanent unchanged current is used
dA: Day
dEL: Delete
dL: Delay between channels A and B
HI: Upper value (high)
Hz: Hertz is the measurement unit for the frequency of impulses per second.
LO: Lower value
mA: Milli-amp is the measurement unit for the electrical current intensity.
MOD: In the modulated mode the frequency and/or the pulse width and/or the intensity
changes in a predened rhythm automatically.
Rd: Ramp down
RU: Ramp up
SEC: Seconds
SY: Synchronous function of the channels A and B
TI: Time
TM: Time of modulation-interval
W/R: Work/Rest mode
µS: Microsecond is the measurement unit for the length of every single impulse.
Program 01 02 03 04 05 06
Mode MOD CONT BURST HAN W/R selectable
Frequency 70-110 Hz 110 Hz 80 Hz 100/2 Hz 50 Hz 1-150 Hz
Pulse Width 200-100 µS 175 µS 150 µS 200/350 µS 300 µS 50-350 µS
Running Time 30 min. 30 min. 30 min. 30 min. 20 min. 1-599 min.
Ramp up 1 sec. 1 sec.
Ramp down 1 sec. 1 sec.
Work Period 10 sec. 2-99 sec.
Rest Period 20 sec. 2-99 sec.