Dolito Bedienungsanleitung / Operators Manual
42
Trouble Shooting Guide
Fault Possible Cause Solution
Trip o - Elec-
trode symbol
ashing, current
won’t increase
above 8 mA)
Electrode lost adhesion to skin.
Check electrodes and replace if
necessary.
Electrode to dry and loses
adhesion to skin.
Replace electrode.
Lead wire not connected
properly.
Check lead wire connections and
ensure they are fully inserted.
Lead wire is faulty. Replace lead wire.
Device is damaged.
Return device to manufacturer for
repair.
Device does not
power on.
Flat battery or no batteries
installed in device.
Check / Replace batteries.
Battery installed in wrong
orientation.
Remove batteries and install accor-
ding to diagrams in battery compart-
ment.
Battery connector damaged or
attened.
Remove batteries, check metal parts
in battery compartment for damage.
Device is damaged. Return to manufacturer for repair.
Device powers
o during use.
Batteries almost at.
Check battery icon during use and
replace batteries if low.
Device is damaged. Return to manufacturer for repair.
Current output is
unstable
Device is in use near a device
emitting strong radio or mag-
netic interference.
Move at least four meters away from
the oending device and resume
treatment.
Treatment
is painful or
uncomfortable
when using skin
electrodes.
Patients skin has cuts or abra-
sions in the area of electrode
placement.
Discontinue use until skin is fully
healed.
Treatment
is painful or
uncomfortable
when using a
probe.
Patient reaction to probe
materials.
Discontinue use and consult a doctor
for an allergy check if discomfort
continues.
Skin becomes
red after treat-
ment.
Device is being used for
periods which are too long for
the patient.
Use device for a shorter treatment
duration.
Treatment is
uncomfortable,
tingling sensa-
tions during use.
Intensity too high for required
treatment.
Reduce intensity setting until treat-
ment is comfortable.
Note: Tingling sensations are normal
and should be felt.