Installation
MTS Landmark™ Tabletop Load Units - Product Information 53
Warning:
Airmounts can be overinflated and then explode.
You can be seriously hurt if an airmount explodes, sending fragments flying.
Only inflate airmounts that have the full weight of the load unit resting on them. Never
inflate an airmount above 0.55 MPa (80 psi). Check airmount pressures with an
accurate pressure gage.
Each airmount's inflated height—measured from the floor to the bottom of the load
unit’s legshould be 83–89 mm (3.25–3.5 in).
6. Gently lower the load unit onto the four isolation pads or airmounts.
A. If your load unit has isolation pads, and the load unit does not sit level, install stock
metal shims between the pads and the floor.
B. If your load unit has airmounts, inflate each airmount in 5 mm (0.20 in) steps to level
the load unit.
C. Check the airmounts’ pressures often to make sure that they are not inflated above:
0.41 MPa (60 psi) for the tabletop load unit.
0.55 MPa (80 psi) for the floor standing load unit.
D. Check the airmounts’ inflated heights to make sure that they are all between:
57 and 63 mm (2.25 and 2.5 in) for the tabletop load unit.
83 and 88 mm (3.25 and 3.5 in) for the floor standing load unit.
E. Adjust the air pressure to level the load unit.