Controller Connections
MTS Exceed™ Series 40 Product Manual Installation
47
Controller Connections
Connecting the Main Power
5 kN through 100 kN
The input voltage of MTS Exceed frames rated 100 kN or less is single phase
200-230 V, 50/60 Hz.
For MTS Exceed frames equal or less than 10 kN, use 3 holes socket for power
input, specification of power wire is H05VVF,3G0.75 mm
2
, compatible with
standards of CCC , CE etc for certification.
For MTS Exceed frames more than 10 kN, use the Single phase 3-wires cable
that is provided for power input, and connect to the customer electrical box and
disconnect. The type of power cable is listed below:
10 kN~100 kN - HO7RN-F 3G2.5 mm
2
,or the same specifications.
compatible with standards of CCC,CE etc for certification.
Note Electrical connections must be made by qualified personnel and is their
responsibility for using the proper power disconnect that conforms to local
codes and regulations when connecting the machine to the building’s main
power.
100 kN above
The input voltage of MTS Exceed frames rated 100kN above is three phases
380
- 415 V or 440 - 480 V
, 50/60 Hz.
The type of power cable is : JZ-602,4G4 mm
2
,or the same specifications.
compatible with the standards of CCC,CE etc for certification.
Electrical disconnect
The customer is responsible for providing an electrical power disconnect that is
easy to operate and easy to reach. It must meet the standards of CCC,CE etc for
certification
Recommended circuit breakers would be ones that are of the thermal magnetic
type with characteristics suitable for large inductive loads(D-type trip
characteristic).if fuses is used, it is recommended that they are of the time-delay
type with dual elements. These recommendations should be followed to avoid
nuisance tripping.
0Exceed product manual.book Page 47 Wednesday, August 27, 2014 2:55 PM

Table of Contents

Other manuals for MTS Systems Exceed 40 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the MTS Systems Exceed 40 Series and is the answer not in the manual?

MTS Systems Exceed 40 Series Specifications

General IconGeneral
BrandMTS Systems
ModelExceed 40 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals