MTS 793 Controller Hardware
Optional Station Configurations
232
RSC Remote Station Control (RSC) is not supported for 6-station
configurations.
8-Station Configuration
The 8-Station configuration can provide eight channels of control.
A typical 8-Channel/8-Station configuration requires 16 Digital
Universal Conditioners and 8 two-stage valve drivers.
HSM power limits HSM power current is limited to 1.5 A per HSM.
Interlocks For 8-station configurations the hwi file must contain the line
INTERLOCKS=8. This line must be a discrete entry, not part of any
other hwi section.
Cross-head interlocks with solenoid power are not supported for 8-
station configurations.
RSC Remote Station Control (RSC) is not supported for 8-station
configurations.
A/D analog inputs Only one set of six A/D analog inputs is supported for 8-station
configurations.