RMX Programmable Power Supplies User Manual
CF46 IP Address Subnet Mask Prefix Setting
If you want to set the IP address manually (MANUAL IP function), set the subnet mask prefix.
This parameter is displayed and can be set only when the LAN interface is in use.
This parameter is applied when you turn the POWER switch off and on or when you execute
APPL with CF40.
CF47 SCPI-RAW Protocol TCP Port Number Setting
Set the TCP port number of the SCPI-RAW protocol for when the LAN interface is in use.
This parameter is applied when you turn the POWER switch off and on or when you execute
CF40. TCP port number 4880 cannot be used.
Parameter Number Value Description
CF46 16 to 31 Sets the subnet mask prefix.
Examples: Display example when “16” is set 255.255.0.0
Display example when “24” is set 255.255.255.0
Display example when “31” is set 255.255.255.254
Value Parameter Description
4000 to 9999
(except 4880, 5024, and 5044)
TCP port number (factory default setting: 5025)

Table of Contents

Other manuals for National Instruments RMX-4123

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the National Instruments RMX-4123 and is the answer not in the manual?

National Instruments RMX-4123 Specifications

General IconGeneral
BrandNational Instruments
ModelRMX-4123
CategoryPower Supply
LanguageEnglish

Related product manuals