EasyManua.ls Logo

National Instruments SCXI Series - Floating AC-Coupled Signal Connection for the SCXI-1120 D; High Common Mode Voltage; Temperature Sensor Connection

National Instruments SCXI Series
73 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 3 Signal Connections
© National Instruments Corporation 3-7 SCXI-1120/D User Manual
The input signal range of an SCXI-1120 input channel is ±5 V/ G
total
referenced to its negative input, where G
total
is equal to the product of the
first-stage and second-stage gains. In addition,
the input channels are
overvoltage protected to 250 V
rms
with power on or off at a maximum of
4.5 mA
rms
sink or source.
SCXI-1120D
When connecting AC-coupled signals to the SCXI-1120D, you do not need
to add the external resistor, as the SCXI-1120D provides a DC path
internally. Refer to Figures 3-6 and 3-7 for further details.
Figure 3-6. AC-Coupled Signal Connection for the SCXI-1120D with
High Common Mode Voltage
Figure 3-7. Floating AC-Coupled Signal Connection for the SCXI-1120D
Temperature Sensor Connection
Pins C2 and C4 are dedicated for connecting to a terminal block
temperature sensor. The temperature sensor is not isolated and is
referenced to the chassis ground. The connection is overvoltage protected
to ±25 VDC with power on and ±15 VDC with power off.
Module
+
-
+
+
-
I
out
V
s
V
+
-
cm
V
High
CMV
511 k
511 k
+
-
+
+
-
I
out
V
s
V
511 k
511 k