30 | ni.com | NI VirtualBench Calibration Procedure
4. Call MSO Configure Analog Channel with the following parameters:
• Enable Channel = 
TRUE
• Channel = mso/1
• Vertical Coupling = DC
• Probe Attenuation = 1x
• Vertical Range = 20
• Vertical Offset = 0
5. Call MSO Run.
6. Call 
MSO Read. Calculate and record the average analog waveform. Verify that this 
measurement falls within the limits in Table 13.
7. Call MSO Stop.
8. Repeat steps 4 to 7 for all of the ranges in Table 13.
Note The 0.02 V range only applies to the VB-8034 and VB-8054.
9. Call MSO Close.
Table 13. Oscilloscope Offset Verification Limits
Range 
(V)
Verification 
Point (V)
1 Year Limits (mV) 24 Hour Limits (mV)
Lower Upper Lower Upper
20 0 -400 400 -100 100
10 -200 200 -50 50
5 -100 100 -30 30
2 -40 40 -12 12
1 -20 20 -6 6
0.5 -10 10 -3 3
0.2 -4 4 -1.2 1.2
0.1 -2 2 -0.8 0.8
0.05 -1 1 -0.6 0.6
0.02
*
-0.4 0.4 -0.3 0.3
* VB-8034 and VB-8054 only.