NI VirtualBench Calibration Procedure | © National Instruments | 65
DC Power Supply Adjustment
This section explains how to adjust the VirtualBench DC power supply. You can choose to 
perform these adjustment procedures with or without performing the verification procedures first. 
Note NI recommends repeating the verification procedures after you perform these 
adjustment procedures. Re-verification ensures that the instrument you have 
calibrated is operating within specifications after adjustments.
Note Before adjustment perform a self-calibration on the DMM performing the 
adjustment.
Note Allow the VirtualBench to warm up for at least 30 minutes before adjusting.
Note Before adjusting the DC power supply, you must enable and set the output 
voltages at full scale for a minimum of 15 minutes.
Adjusting Voltage Programming and Measurement Accuracy
Complete the following steps to adjust voltage programming and measurement accuracy of 
VirtualBench.
1. Reset the DMM.
2. Call 
PS Initialize Calibration with the following parameters:
• Cal Type = 
Voltage 
• Password = NI (default)
3. Connect Channel +6 V to Volts HI, and GND to Volts LO as shown in Figure 35.