NI VirtualBench Calibration Procedure | © National Instruments | 33
• Probe Attenuation = 1x
• Vertical Range = 20
• Vertical Offset = 0
Note
Configure the oscilloscope calibrator for high impedance loads.
5. Set the oscilloscope calibrator DC voltage output to -90% of the range, or -18 V. 
6. Wait at least 5 seconds for the oscilloscope calibrator to settle.
7. Call 
MSO Run.
8. Call 
MSO Read. Calculate and record the average analog waveform. Save this value as 
V
meas_neg
. 
9. Set the oscilloscope calibrator DC voltage output to +90% of the range, or +18 V.
10. Wait at least 5 seconds for the oscilloscope calibrator to settle.
11. Call 
MSO Run.
12. Call MSO Read. Calculate and record the average analog waveform. Save this value as 
V
meas_pos
.
13. Call 
MSO Stop.
14. Calculate the gain error as follows:
Where V
pos
 and V
neg
 are listed in Table 14.
Verify that this calculation falls within the limits in Table 14.
Table 14.  Oscilloscope Gain Verification Limits
Range (V) V
pos
 (V) V
neg
 (V)
1 Year Limits (%) 24 Hour Limits (%)
Lower Upper Lower Upper
20 18 -18 -2 2 -0.8 0.8
10 9 -9 -2 2 -0.8 0.8
5 4.5 -4.5 -2 2 -0.8 0.8
2 1.8 -1.8 -2 2 -0.8 0.8
1 0.9 -0.9 -2 2 -0.8 0.8
0.5 0.45 -0.45 -2 2 -0.9 0.9
0.2 0.18 -0.18 -2 2 -0.9 0.9
0.1 0.09 -0.09 -2 2 -1 1
V
meas_pos
V
meas_neg
–()
V
pos
V
neg
–()
-------------------------------------------------------- - 1– 100%×