NI VirtualBench Calibration Procedure | © National Instruments | 35
18. Repeat steps 2 to 16 for all ranges on channel 2.
Note When calling MSO Configure Analog Channel in step 4, change the
Channel input to mso/2.
19. Remove the connections from channel 2.
20.
(VB-8034/8054) Connect VirtualBench channel 3 to the oscilloscope calibrator active
head, as shown in Figure 17.
Figure 17. VB-8034/8054 Oscilloscope Gain Channel 3 Connection Diagram
21. (VB-8034/8054) Repeat steps 2 to 16 for all ranges on channel 3.
Note (VB-8034/8054) When calling MSO Configure Analog Channel in
step 4, change the Channel input to mso/3.
1 VirtualBench Oscilloscope 2 Oscilloscope Calibrator 3 Oscilloscope Calibrator Active Head