Revised 11/21/18 4-3
Electromagnetic Compatibility (EMC)
Nicolet Cortical Stimulator
The Natus Neurology Incorporated Nicolet Cortical Stimulator needs special
precautions regarding EMC and needs to be installed and put into service
according to the Electromagnetic Compatibility (EMC) information provided in
the following Manufacturer’s Declaration.
Portable and mobile Radio Frequency (RF) communications equipment can
affect the operation of the Nicolet Cortical Stimulator.
The following is a list of the cables and other accessories that are used as part of
the Nicolet Cortical Stimulator that comply with sections 201.17 and 202 of IEC
60601-2-40, Ed. 2.0:
• SSU Interface Cable - 5m (Control Unit to SSU)
• SSU Interface Cable - 1m (1st to 2nd SSU)
• Serial Interface Cable - 5m (Control Unit to Computer)
Use of cables or accessories other than those specified, with the exception of
cables and accessories sold by the manufacturer Natus Neurology Incorporated
as replacement parts for internal components, may result in increased
EMISSIONS or decreased IMMUNITY of the Nicolet Cortical Stimulator.
This Nicolet Cortical Stimulator is intended for use by healthcare
professionals only. This system may cause radio interference or may disrupt
the operation of nearby equipment. It may be necessary to take mitigation
measures, such as reorienting the Nicolet Cortical Stimulator or shielding
the location.