2. Safety and Standards Conformity
2.1. Safety and Standards Conformity for Natus Trex HD Monitoring
2.1.1. Standards of Compliance and Normative References
Electroencephalograph system, model Trex HD Monitoring; rated two AA 1.5V batteries, 0.065A,
0.170W.
1. Type of protection against electric shock: Internally Powered Equipment
2. Degree of protection against electric shock: Type BF
3. Degree of protection against ingress of water: IPX0
4. Mode of operation: Continuous
5. Environmental Conditions: Normal: 5-40°C, 15-93% rH, 700-1060hPa
The essential performance of the Trex HD Monitoring amplifier system is defined in the particular
standards IEC 60601-2-26 and ISO 80601-2-61. It is EEG equipment that incorporates a pulse oximeter
without an alarm.
The Trex HD Monitoring and its accessories have been designed to comply with the following national
and international standards.