What to do if natus Xltek EMU40EX Amplifier system not responding?
- BbookermelissaAug 14, 2025
If the Natus Amplifier system is not responding or functioning correctly, turn the Base Unit off and then on again. This will reset the system.
What to do if natus Xltek EMU40EX Amplifier system not responding?
If the Natus Amplifier system is not responding or functioning correctly, turn the Base Unit off and then on again. This will reset the system.
Why is there interference from other devices with my natus Amplifier?
If you're experiencing interference with your Natus Amplifier, it might be due to other devices on the same circuit. Unplug devices like printers, mechanical beds, vacuum cleaners, or other potential sources of interference.
What to do if cables are not connected properly between the EMU40EX and the computer for natus Amplifier?
If the cables are not connected properly between the EMU40EX and the computer, make sure all cables are correctly connected. If using a LAN, ensure that both the hardware and software are configured correctly.
How to fix improperly grounded electrical system for natus Xltek EMU40EX?
To address an improperly grounded electrical system with your Natus Amplifier, install a medical-grade ground to ensure your clinic has a properly grounded electrical system.
What to do if patient electrodes are not connected to the correct channels in the Breakout Box for natus Amplifier?
If the patient electrodes are not connected to the correct channels in the Breakout Box, ensure that they are connected to the correct channels.
What to do if electrodes touching and causing a short circuit in natus Amplifier?
If any of the electrodes are touching, they are causing a short circuit and will develop an artifact. Ensure that electrodes are not touching.
What to do if patient electrodes fit improperly into the Breakout Box for natus Xltek EMU40EX Amplifier?
If the patient electrodes fit improperly into the Breakout Box, make sure that they fit properly and are not loosely connected.
How to adjust gain and timebase settings on natus Xltek EMU40EX Amplifier?
To adjust the gain and timebase settings on the Natus Amplifier, check the gain and timebase settings to ensure they are appropriate for the current test. You may also want to check the LFF, HFF, and Notch Filter settings. Choose Edit > Settings > Montage. Right-click a cell in the appropriate column and select a value.
Lists the main components of the EMU40EX system.
Defines critical performance parameters for the amplifier.
Lists relevant safety and performance standards the system complies with.
Lists conditions under which the system should not be used.
Provides essential safety precautions for general operation and maintenance.
Outlines precautions to prevent damage from static electricity.
Details the analog input and channel specifications of the amplifier.
Details the digital sampling and resolution specifications.
Explains how to connect the EMU40EX system to a computer.
Provides initial steps and notes for beginning to use the system.
Details the correct procedure for safely shutting down the system.
Explains how data is transferred between the Breakout Box and Base Unit.
Describes how to verify the amplifier's factory calibration.
Explains how to perform a channel test to check signal processing.
Details how to perform an impedance check for electrode contact quality.
Explains the underlying principles of the Masimo SET pulse oximeter.
Details the specifications for the XLTEK Trolley.
Offers specific advice for cleaning and maintaining the amplifier.
Provides a step-by-step checklist for diagnosing issues.
Provides instructions and contact information for Natus Technical Support.