EMU40EX™ User and Service Manual
40
Amplifier Usage and Features
Getting Started
NOTE: In the event of a power failure, the current recording will resume using the last
programmed settings upon the restoration of power.
Placement of the Operator and Patient
It is expected that the operator of the system will stand or sit in front of the computer, but not
continuously. The patient is typically lying in a bed located beside the system cart or amplifier and is in
no way supported by the equipment.
Amplifier units are patient-worn. Refer to the section Adding the EMU40EX Breakout Box to a Harness
or Belt for details on the placement of the Breakout Box on the patient.
At no point should the system be leaned against or rested upon. Refer to the Transport System
Specifications and Maintenance section for placement, details, and cautions for the different cart
transportation options.
Refer to the corresponding Instructions for Use for all system components prior to use. This should
include, but is not limited to: cameras, computers, stimulators, and software.
Beginning a study
Once the equipment has been installed by your Natus qualified representative and a patient has been
connected to the EMU40EX system, a new EEG study can be started. For details on beginning a new
EEG study, consult the NeuroWorks manual.
NOTE: In order to record a study the Base Unit must be connected to a PC and the
Base Unit must be connected to the Breakout Box as described in EMU40EX
Breakout Box to Base Unit Data Link.
Powering Down the System
Utilize the following steps to ensure your system is powered down completely and safely.
1. Close any active studies in the NeuroWorks software.
2. Shut the computer down; ensuring to follow the proper shut down procedure.
3. On the base LCD press and hold the Shutdown button for 3 seconds.
4. Unplug the power cord from the wall.Adding the EMU40EX Breakout Box to a belt or a lanyard
Placing the EMU40EX Breakout Box on the patient can be accomplished by using either the belt or the
lanyard setup.
The pouch and belt or lanyard system must be worn over top of the patient’s clothes or
hospital garment.