9. Appendicies
Appendix 1: Specifications
Power Required
110V AC ± 20%, 45 - 65 Hz (US only)
220V AC ± 20%, 45 - 65 Hz (Europe only)
240V AC ± 20%, 45 - 65 Hz (UK only)
100V - 240V AC ± 20%, 45 - 65 Hz (with in-line lead)
Dimensions
Depth - 160 mm (6.4 inches)
Width - 215 mm (8.5 inches)
Height - 110 mm (4.25 inches)
LCD Screen Dimensions
Width – 88 mm (3.5 inches)
Height – 44 mm (1.3 inches)
Adjustable LCD Screen: up to 45 degree incline
Weight
Pro Advantage Urine Analyzer (without power supply) – 2.36 kg (5.21 lb)
Operating Temperature
18°C to 30°C (64°F to 86°F)
Operating Humidity
18% to 80% Relative Humidity
Optimal Operating Temperature
22°C to 26°C (72°F to 79°F)
Optimal Operating Humidity
35% to 55% Relative Humidity
Optimal ranges ensure that the reagent results are optimized for performance and stability. At temperatures under 22°C
(72°F), urobilinogen and leukocyte results may be decreased, and at temperatures above 26°C (79°F), increased.
Altitude: 2000 m (6562 ft)
Instrument Memory
1000 Patient test results (With Patient Identification)