SECTION 2
SERHT50NA A0808 2-7
41. Factory Default Parameters
P
atient Settings:
MODE A/CMV PEEP/CPAP 0 cmH
2
O
Volume Control 500 mL Psupport 0 cmH
2
O
t
I
1.0 sec Humidifier Off
ƒ 15 b/min Buzzer Volume Loud
Ptrig –1.0 cmH
2
O
Paw Alarms 5 cmH
2
O ▼Paw
40 cmH
2
O ▲Paw
V
•
I
Alarms 3 L/min ▼V
•
I
20 L/min ▲V
•
I
User Set Up:
Power Save On
Pressure Units cmH
2
O
Auto Panel Lock Disabled
Set Up User
42. Patient Circuit Reusable 22 mm I.D. adult/pediatric circuit with 3/16 inch
(4.8
mm) I.D. proximal pressure sensing line, 1/8 inch (3.2 mm)
I.D. exhalation valve control drive line, and exhalation valve.
43. Exhalation Valve N
EWPORT'S HT50 exhalation valve (P/N HT600039) is
manufactured and designed specifically for the N
EWPORT HT50
Ventilator. N
EWPORT MEDICAL does not approve of the use of any
type or brand of exhalation valve that has not been tested and
approved by N
EWPORT MEDICAL for use with the HT50.
HT50-H, HT50-HB Humidifier Specifications
(operates on A.C. power only)
Set Target T
emperature Range: 19ºC to 39ºC
Operating Water V
olume: 300 mL
Usable Volume of
W
ater Bottle: 265 mL
Compliance at Minimum
W
ater Level (Refill Line): 0.5 mL/cmH
2
O / mbar @ 23ºC
Compliance at Maximum
Water Level (Full Line): 0.33 mL/cmH
2
O / mbar @ 23ºC
Intended Use: Adult and pediatric patients whose supraglottic airway is or is
not b
ypassed.