SPECIFICATIONS
2-8 SERHT50NA A0808
Warm-Up Time: 30 minutes
Gas Leakage: 2 mL/min at airway pressure of 80 cmH
2
O / mbar
Humidifier Output: 33.8 mg/L at a continuous flow of 10 L/min @ 39ºC
Maximum Operating
Airwa
y Pressure: 100 cmH
2
O / 98 mbar
Maximum Temperature at
the P
atient Wye That
Triggers an Alarm: 41ºC
(optional) Air / Oxyg
en Entrainment Mixer Specifications
Pneumatic Requirements:
Oxygen 35 to 90 psig (2.4 to 6.2 Bar) full operating range
40 to 70 psig (2.7 to 4.8 Bar) accuracy
} .08
Air Atmospheric pressure
F
I
O
2
Control: adjusted continuously from 0.21 to 1.00
WARNING Continuous oxygen monitoring is required for patient
safety. The HT50 does not have a built-in alarm system to notify user
of a failure or disconnect of the oxygen source.
(optional) Oxygen Blending Bag Kit Specifications
Pneumatic Requirements:
Oxygen 0-10 L/min (calibrated)
Air Atmospheric pressure
F
I
O
2
Control: F
I
O
2
, indirectly adjusted from 0.21 up to 1.00
via oxygen flow (L/min)
WARNING Continuous oxygen monitoring is required for patient
saf
ety. The HT50 does not have a built-in alarm system to notify user
of a failure or disconnect of the oxygen source.