Quick Reference Guide:
Power OFF:
1. Disconnect Battery
a. Disconnect (+) Positive Cable (Red)
b. Disconnect (-) Negative Cable (Black)
c. Verify Battery Reverse Indicator
2. Unplug NexPump’s AC Power Cord
Power ON:
1. Plug in NexPumps AC Power Cord
2. Connect Battery
a. Connect (-) Negative Cable (Black)
b. Connect (+) Positive Cable (Red)
c. Verify Battery Reverse Indicator
Override Mode:
1. Set Override switch to up position
2. Both Pumps should turn on.
3. Battery Life will not be calculated
4. Clog or Vapor lock detection is disabled.
5. Pumps will turn off for 5 seconds every minute to clear
possible Vapor Lock Conditions.
6. Set Override switch to down position to turn pumps off and
return to normal.
Clear Alarm(s):
1. Press Function Button Until ‘C’ appears (1 second).
2. Release Function Button.
3. Display shows ‘Alarm Cleared’ if there are any that can be
cleared, otherwise displays ‘No CLR Alarms’.
Silence Alarm:
1. Press Function Button Until ‘S’ appears (3 seconds).
2. Release Function Button.
3. Display shows ‘Alarm Silenced’
Note: Will be silent for 24 hours and re-enable itself automati-
cally
UnSilence Alarm:
1. Press Function Button Until ‘U’ appears (3 seconds). (A
‘U’ will only appear if the alarm is silenced already)
2. Release Function Button.
3. Display shows ‘Alarm UnSilenced’
Manual Test:
1. Press Function Button Until ‘T’ appears (6 seconds).
2. Release Function Button.
3. Test is scheduled.
Get Serial Number:
1. The NexPumps serial number is located on the rear unit or
proceed to step 2.
2. Press Function Button Until ‘N’ appears (9 seconds).
3. Release Function Button.
4. Display shows SN-NPAxxxxxxxx where ‘xxxxxxxx’
is the serial number
Get Firmware Revision:
1. Press Function Button Until ‘A appears (12 seconds).
2. Release Function Button.
3. Display shows revision number in the following format
‘x.xx.xxxx oooo’. The ‘x.xx.xxxx’ is the revision number and
the ‘oooo’ is any options installed.
Phone Monitor:
1. Press Function Button Until ‘M’ appears (24 seconds).
2. Display shows ‘Ph-Monitor ? Y N’.
3. Depress and hold Function Button, display will toggle
between ‘Y’ and ‘N’, release when at correct setting.
Battery Reset:
1. Press Function Button Until ‘R’ appears (18 seconds).
2. Release Function Button.
3. Display shows ‘B-Rst ? xxx AH’.
4. Enter Battery Amp Hours
5. Battery Charge will start.
Stats Information:
1. Press Function Button Until ‘I’ appears (21 seconds).
2. Release Function Button.
3. Display will toggle through NexPump statistics.
Disable Alarm:
1. Press Function Button Until ‘D’ appears (27 seconds).
2. Display shows ‘Alarm Disabled’.
3. Alarm remains disabled for 5 days then automatically re-
enabled
Enable Alarm:
1. Press Function Button Until ‘E’ appears (27 seconds).
2. Display shows ‘Alarm Enabled’.
Check IMPOR
TANT
INFORMATION DURING AN
EVENT on
page 17