Chapter 1 Microscopy Procedures
23
Microscopy Procedures
Sensitive Tint Plate Microscopy
Chapter 1-4
5
Place a specimen on the stage, and move the
stage to bring the target into view.
6
Remove the analyzer and the polarizer from the
optical path.
Pull out the analyzer IN/OUT knob from the sensitive tint
plate analyzer unit to remove the analyzer from the optical
path.
The polarizer unit for first-order red compensation has not
been installed at this point.
Removing the analyzer from the optical path
7
Focus on the specimen. (See Chapter 2 “2 Focusing on the Specimen (Vertical Stage Movement)” fo
r
details)
8
Adjust the diopter. (See Chapter 2, “4 Adjusting the Diopter” for details)
9
Adjust the interpupillary distance.
10
Focus and center the condenser. (See Chapter 2, “5 Focusing and Centering the Condenser” for details)