65
V. External Communications Control
[r] [VER] [<CR>]
Reads the program version number.
When properly finished [a][VER][data][<CR>]
When control error occurred [n][VER][error code][<CR>]
[data]: V*.** (* denotes numeral. Example: V1.00)
[r] [PNM] [<CR>]
Reads the program name. You can identify devices connected on the communication line from
an external device.
When properly finished [a][PNM][data][<CR>]
When control error occurred [n][PNM][error code][<CR>]
[data]: LV150A
[c] [SAS] [data] [<CR>]
Sets the status output setting for rotation of the nosepiece.
When properly finished [o][SAS][<CR>]
When control error occurred [n][SAS][error code][<CR>]
[data]: 0 (status output disabled), 1 (status output enabled), or 2 (external light source shutter
control enabled)
[r] [SAR] [<CR>]
Reads the status output setting (the value set with cSAS command).
When properly finished [a][SAR][data][<CR>]
When control error occurred [n][SAR][error code][<CR>]
[data]: 0 (status output disabled), 1 (status output enabled), or 2 (external light source shutter
control enabled)
[s] [SAE] [data] [<CR>]
Outputs the status with the rotation of the nosepiece when the status output is enabled. No
response is required from the external device to this unit.
Status output is [s][SAE][data][<CR>]
[data]: 0 (at the start of nosepiece rotation), or 1 to 5 (address) (at the end of the nosepiece
rotation)
[c] [DEF] [<CR>]
Initializes the control data to the factory default.
When properly finished [o][DEF][<CR>]
When an error occurred [n][DEF][error code][<CR>]
NOTE: Please refer to “5. Communication Formats” for description of [error code].

Table of Contents