4.3 Setting Up the Microscope
When you have finished attaching the optical parts to the microscope, you must register information about the
objectives mounted on the microscope.
(1) Turn off the microscope.
Disconnect the AC adapter and turn off the
microscope.
(2) Turn on the microscope while pressing the
brightness control knob and illumination on/off
button.
The SET UP screen appears.
Information on the objective
The address of the nosepiece that is in the optical
path and information on the the registered
objective such as the objective name and
magnification are displayed in the OBJECTIVE
field.
Browse and select items on the SET UP screen by
using the brightness control knob and illumination
on/off button. To move the cursor, turn the brightness
control knob. To select an item, press the illumination
on/off button.
Turning on the microscope while pressing the brightness
knob and illumination on/off button
SET UP screen
Browse items Select
/values
(3) Input information about the objectives in each
position of the nosepiece.
1. Move the cursor () to “OBJECTIVE” and
press the illumination on/off button.
The cursor moves to the displayed name and
magnification (for example, [E Plan / Mag: 4x]).
2. Turn the nosepiece to bring the address to
be registered into the optical path.
When you turn nosepiece, the registered name
and magnification of the new objective is
displayed. If you purchased the microscope as a
set, the objectives included in the set have
already been pre-registered.
3. Turn the brightness control knob to specify
the name and magnification of the objective.
Information on standard objectives is registered
in the microscope.
If you are using a non-standard objective, select
an item where only the magnification is
displayed.
For positions where no objective is mounted,
specify “None.”
4. Repeat steps 2 and 3 to register information
about all the desired positions.
5. Press the illumination on/off button.
The cursor moves to “ECO TIME.”
1. Select OBJECTIVE.
2. Turn the nosepiece to select the address.
3. Select the objective using the brightness control knob.
Objective setting values
Objective type
Magnification
value
Objective type
Magnification
value
1 E Plan 4x 10 1x
2 E Plan 10x 11 2x
3 E Plan 40x 12 4x
4 E Plan 60x 13 10x
5 E Plan Oil 100x 14 20x
6 Achro Ph1 10x 15 40x
7 Achro LWD Ph1 20x 16 50x
8 Achro Ph2 40x 17 60x
9 Achro Oil Oh3 100x 18 100x
19 None