Chapter 9 Troubleshooting
133
3
Motorized Devices
Problem Cause Measure
The product cannot be
turned on.
The power cable is not connected. Connect the power cable properly.
Operating the remote
controller does not cause
a motorized device to
operate.
The power of the P2-CTLA Control Box is not
turned on or the AC adapter is not connected.
Check the connections between the AC adapter
and the P2-CTLA Control Box, and then move the
power switch to the “|” position to turn on the
power.
The devices are not connected or the
connections are not made correctly.
Connect the devices correctly.
The motorized focus unit
does not operate.
The unit is placed in escape status. Exit the escape status.
The STOP switch is pressed.
Press the STOP switch to release the forced stop
state.
The motorized focus unit
does not move down
completely.
Limit setting has been made. Cancel the limit setting.
The motorized zoom
function works
unexpectedly.
The ALZ function is enabled. Disable the ALZ function.
The LCD of the remote
controller is invisible.
The backlight is turned off. Turn the backlight control dial.
The LED of the remote
controller is off.
The backlight control dial is set to the
minimum level.
Turn the backlight control dial.
The light of the LED
diascopic illumination base
cannot be adjusted.
The LED power ON/OFF switch is set to OFF. Set the LED power ON/OFF switch to ON.
The control right has not been matched. Match the control right.

Table of Contents