Chapter 7 Specific Operations
68
1
Turning On the Power
This section lists the devices of the microscope system that require a power supply.
Connect the device to the relevant power source before turning on the power. For details on the power supply required by
each device, see Chapter 8, “Functions and Operations of the Devices” or the instruction manual supplied with the relevant
device.
P2-CTLA Control Box
P2-CTLB Control Box
P2-DBL LED Diascopic Illumination Base, P-DSL32 LED Diascopic Illumination Stand
Power supply (control box/12 V AC adapter)
Power to the P2-DBL LED Diascopic Illumination Base and P-DSL32 LED Diascopic Illumination Stand is supplied by
connecting a control box or a 12 V AC adapter. (The P-DSL32 can only use the P2-CTLB Control Box.) To provide
diascopic illumination without using a control box, connect a 12 V AC adapter.
LED power on/off and illumination control
Turn on or off the LED diascopic illuminator using the LED power ON/OFF switch on the P2-DBL LED Diascopic
Illumination Base or P-DSL32 LED Diascopic Illumination Stand.
Adjust illumination using the LED brightness control dial on the base (stand). To enable adjustment using the LED
brightness control dial on the base (stand), the LED brightness control selection switch must be turned off.
When a control box is connected to the base (stand), turn on the LED brightness control selection switch. In this state,
diascopic illumination can be turned on or off, brightness can be controlled, and light intensity can be monitored from the
control box. Note that illumination cannot be controlled from the control box, even if the LED brightness control selection
switch is turned on, unless the LED power ON/OFF switch on the base (stand) is turned on.
P2-DBF Fiber Diascopic Illumination Base, P-DSF32 Fiber Diascopic Illumination Stand (Refer to the instruction manual
supplied with the C-FLED2 LED Light source for fiber illuminator.)
P2-CIA Coaxial Epi Illuminator (Refer to the instruction manual supplied with the C-FLED2 LED Light source for fiber
illuminator.)
P2-EFLI Epi Fluorescence Attachment, P2-EFLM Motorized Epi Fluorescence Attachment (Refer to the instruction
manual supplied with the C-HGFI/C-HGFIE HG Precentered Fiber Illuminator.)
C-FDF Flexible Double Arm Fiber Illumination Unit (Refer to the instruction manual supplied with the C-FLED2 LED Light
source for fiber illuminator.)
C-FID2 Double Arm Fiber Illuminator (Refer to the instruction manual supplied with the C-FLED2 LED Light source for
fiber illuminator.)
P2-FIR Ring Fiber Illumination Unit (Refer to the instruction manual supplied with the C-FLED2 LED Light source for fiber
illuminator.)
P2-FIRL2 LED Ring Illumination Unit 2 (Refer to the instruction manual supplied with the P2-FIRL2 LED Ring Illumination
Unit 2.)
P-DF2 LED Dark Field Unit 2 (Refer to the instruction manual supplied with the P-DF2 LED Dark Field Unit 2.)
C-LSL2 LED Epi Illuminator 2 (Refer to the instruction manual supplied with the C-LSL2 LED Epi Illuminator 2.)
When using an illuminator whose light source is an HG precentered fiber illuminator or fiber illumination LED light source,
also turn on the power to the light source. (Refer to the instruction manual supplied with the light source.)
When using multiple power supplies, they need not be turned on in a particular order except for the following cases:
When using a camera control unit to capture images, power on the camera control unit at the end.
When controlling a C-HGFIE HG Precentered Fiber Illuminator from NIS-Elements, etc. via a control box, power on the
HG precentered fiber illuminator before the control box.
Press the power switch to the “|” position to turn the power on and to the “O” position to turn the power off.
When power is turned on, the power LED lights.