10. Compare result to Control range printed on
Strip vial label. If result is in range, System can
be used for testing blood.If result does not fall
within range, repeat test using a new Strip.
Note:
Control Test result shows the Control Symbol
in the Display.
If Control Test result is outside range,
test again. If result is still outside range,
System should not be used for testing
blood. Call for assistance (see front
cover for phone number).
11. Remove Strip from Meter and discard. Meter
turns off. Recap Control bottle tightly.