Running Glucose Control Solution
Important Information for Glucose Control Solution
• UseonlyNovaMaxGlucoseControlSolutionsforthetest.
• Checktheexpirationdateonthecontrolsolutionvial.Donot
use control solution past the expiration date or you may get
inaccurate results.
• Storeonlyfor3monthsafterrstopening.Whenyouopen
a new vial of control solution, count forward 3 months and
write that date on the label of the control solution vial. Dis-
card any remaining solution after the date you have written
on the vial.
• Storethecontrolsolutiontightlyclosedatroomtemperature
below86°F(30°C).Donotrefrigerateorfreeze.
• Shakethecontrolsolutionwellbeforeusing.
Caution: The Nova Max Glucose Control Solution range printed on
the glucose test strip vial is for control solution only. It is used to test
the performance of the monitor and test strip. It is not a recom-
mended range for your blood glucose level.
If your control solution test results continue to fall outside the
range printed on the test strip vial:
• TheNovaMaxPlusMonitormaynotbeworkingproperly.
• Donotusethemonitortotestyourblood.
• InsidetheUSA,callCustomerServiceat1-800-681-7390.
Outside the USA, contact your Nova supplier.
Nova Max Plus Monitor OG PN46171D1 EN.indd 14 9/5/12 3:22 PM

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Nova Max PLUS and is the answer not in the manual?

Nova Max PLUS Specifications

General IconGeneral
Test Time5 seconds
HCT Range30 - 55%
Operating Temperature50-104°F (10-40°C)
Operating Humidity10-90% RH
Blood Sample Size0.5 µL
Measuring Range20-600 mg/dL