Testing a Glucose Quality Control Solution
1. Insert a glucose test strip into the monitor. If monitor was o,
the screen displays all segments for 2 seconds then the blink-
ing blood drop symbol and “Glu appear.
NOTE: If the strip is removed before you start the test, the screen
goes blank.
2. Press the left/right
buttons to indicate this
sample is a control. (CTL is shown on the display.)
NOTE: It is important to select control solution test so the test result
does not appear to be one of your blood test results.
NOTE: If a test is not performed within 2
minutes from the insertion of the test strip, the
screen goes blank. To perform a test, take out
then replace the test strip starting from Step 1.
3. Shake the control solution vial. Discard
adropbeforeuse.Squeezeadropof
control solution onto a clean, hard, dry
surface, i.e., control cap.
4. Pick up the monitor with test strip inserted
and touch the test strip to the control solu-
tion drop.
Running Glucose Control Solution
Nova Max Plus Monitor OG PN46171D1 EN.indd 15 9/5/12 3:22 PM

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Nova Max PLUS and is the answer not in the manual?

Nova Max PLUS Specifications

General IconGeneral
Test Time5 seconds
HCT Range30 - 55%
Operating Temperature50-104°F (10-40°C)
Operating Humidity10-90% RH
Blood Sample Size0.5 µL
Measuring Range20-600 mg/dL

Related product manuals