Under appropriate lighting (500 lx or higher), inspect the Motor and the Motor cord for
contamination.
If any visible contamination remains, repeat the process until it is visually clean. Be sure to repeat
from the procedure No.4.
5
Post-use Maintenance
≥ 500 lx
< Disinfection >
Wipe the surface of the Motor, the Motor Cord and the Holder of the Control Unit with the
disinfectant wipes (Minuten Wipes: ALPRO).
1
Disconnect the Motor from the Motor Cord.
Motor Cord: Place the Holder of the Control Unit.
Motor: Proceed to section "6-4 Sterilization"
2
Minuten Wipes (ALPRO)